Liquid Particle Counter

Specifications:

  • Sample input: Low pressure: 0 to 6 bar – batch and online mode 
  • Volumetric particle sensor PAMAS HCB-LD-50/50: Maximum particle concentration: 24,000 particles per ml at a coincidence rate of 7.8 % 
  • Nominal flow rate: 25 ml/min 
  • Calibration according to common cleanliness standards and based on international calibration standards which are traceable to the NIST (National Institute of Standards and Technology).
  • Up to 32 free adjustable size channels may be selected and stored in the system. 3 different channel settings for polyglycol measurements are pre-set in the system:
  • SAE AS 4059, channels A – F, 4 – 70 µm(c): 

Calibration in six size channels: > 4, > 6, > 14, > 21, > 38, > 70 µm(c)

  • SAE AS 4059, channels B – F, 6 – 70 µm(c):

Calibration in five size channels: > 6, > 14, > 21, > 38, > 70 µm(c)

  • NAS 1638, 5 – 100µm: 

Calibration in the size channels > 5, > 15, > 25, > 50, > 100 µm